The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial
作者: Sergius A. R. de Oliveira; Lorena M. C. Lucio; Norma S. P. Modolo; Yoko Hayashi; Mariana G. Braz; Lídia R. de Carvalho; Leandro G. Braz; José Reinaldo C. Braz
刊名: PLOS ONE
ISSN: 1932-6203
出版年: 2017
发表日期: 2017-1-27
卷: 12, 期:1
语种: 英语
英文关键词: Gases ; Anesthesia ; Breathing ; Humidity ; General inhalational anesthesia ; Intubation ; Surgical and invasive medical procedures ; Anesthetics
英文摘要: Background During anesthesia, as compared with intensive care, the time of the tracheal intubation is much shorter. An inhaled gas minimum humidity of 20 mgH2O.L-1 is recommended to reduce the deleterious effects of dry gas on the airways during anesthesia with tracheal intubation. The Fabius GS Premium® anesthesia workstation (Dräger Medical, Lübeck, Germany) has a built-in hotplate to heat gases in the breathing circuit. A heat and moisture exchanger (HME) is used to further heat and humidify the inhaled gas. The humidity of the gases in the breathing circuit is influenced by the ambient temperature. We compared the humidity of the inhaled gases from a low-flow Fabius anesthesia workstation with or without thermal insulation (TI) of the breathing circuit and with or without an HME. Methods We conducted a prospective randomized trial in 41 adult female patients who underwent elective abdominal surgery. The patients were allocated into four groups according to the devices used to ventilate their lungs using a Dräger Fabius anesthesia workstation with a low gas flow (1 L.min-1): control, with TI, with an HME or with TI and an HME (TIHME). The mean temperature and humidity of the inhaled gases were measured during 2-h after connecting the patients to the breathing circuit. Results The mean inhaled gas temperature and absolute humidity were higher in the HME (29.2±1.3°C; 28.1±2.3 mgH2O·L-1) and TIHME (30.1±1.2°C; 29.4±2.0 mgH2O·L-1) groups compared with the control (27.5±1.0°C; 25.0±1.8 mgH2O·L-1) and TI (27.2±1.1°C; 24.9±1.8 mgH2O·L-1) groups (P = 0.003 and P<0.001, respectively). Conclusions The low-flow Fabius GS Premium breathing circuit provides the minimum humidity level of inhaled gases to avoid damage to the tracheobronchial epithelia during anesthesia. TI of the breathing circuit does not increase the humidity of the inhaled gases, whereas inserting an HME increases the moisture of the inhaled gases closer to physiological values.
URL: http://journals.plos.org/plosone/article/file?id=10.1371/journal.pone.0170723&type=printable
Citation statistics:
资源类型: 期刊论文
标识符: http://119.78.100.158/handle/2HF3EXSE/25934
Appears in Collections:过去全球变化的重建
影响、适应和脆弱性
科学计划与规划
气候变化与战略
全球变化的国际研究计划
气候减缓与适应
气候变化事实与影响

Files in This Item: Download All
File Name/ File Size Content Type Version Access License
journal.pone.0170723.pdf(968KB)期刊论文作者接受稿开放获取View Download

作者单位: Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Biostatistics, Institute of Biosciences, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil;Department of Anesthesiology, Botucatu Medical School, UNESP–Univ Estadual Paulista, Botucatu, Sao Paulo State, Brazil

Recommended Citation:
Sergius A. R. de Oliveira,Lorena M. C. Lucio,Norma S. P. Modolo,et al. The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial[J]. PLOS ONE,2017-01-01,12(1)
Service
Recommend this item
Sava as my favorate item
Show this item's statistics
Export Endnote File
Google Scholar
Similar articles in Google Scholar
[Sergius A. R. de Oliveira]'s Articles
[Lorena M. C. Lucio]'s Articles
[Norma S. P. Modolo]'s Articles
百度学术
Similar articles in Baidu Scholar
[Sergius A. R. de Oliveira]'s Articles
[Lorena M. C. Lucio]'s Articles
[Norma S. P. Modolo]'s Articles
CSDL cross search
Similar articles in CSDL Cross Search
[Sergius A. R. de Oliveira]‘s Articles
[Lorena M. C. Lucio]‘s Articles
[Norma S. P. Modolo]‘s Articles
Related Copyright Policies
Null
收藏/分享
文件名: journal.pone.0170723.pdf
格式: Adobe PDF
此文件暂不支持浏览
所有评论 (0)
暂无评论
 

Items in IR are protected by copyright, with all rights reserved, unless otherwise indicated.